Living up to LifeLeica Angle Two™Computer-Assisted Stereotaxic SystemUser Manual
ComponentsLeica Angle Two™ User Manual Rev A02 © Leica Biosystems Richmond Inc. 2010 103.1 U FrameFigure 4 shows the Angle Two U frame with a rat ada
ComponentsLeica Angle Two™ User Manual Rev A02 © Leica Biosystems Richmond Inc. 2010 11Caution Do not use the rat incisor bar on a mouse. A mouse ada
ComponentsLeica Angle Two™ User Manual Rev A02 © Leica Biosystems Richmond Inc. 2010 12The ear bars are locked into position using the locking levers
ComponentsLeica Angle Two™ User Manual Rev A02 © Leica Biosystems Richmond Inc. 2010 133.2 Manipulator: Linear MovementsThe manipulator is attached t
ComponentsLeica Angle Two™ User Manual Rev A02 © Leica Biosystems Richmond Inc. 2010 14Each linear movement is controlled by a coarse drive and, in t
ComponentsLeica Angle Two™ User Manual Rev A02 © Leica Biosystems Richmond Inc. 2010 15Fine DriveThe fine drive (see Figure 9) provides greater contr
ComponentsLeica Angle Two™ User Manual Rev A02 © Leica Biosystems Richmond Inc. 2010 163.3 Manipulator: Tilt, Rotation and Swing-OutTilt, rotation an
ComponentsLeica Angle Two™ User Manual Rev A02 © Leica Biosystems Richmond Inc. 2010 17TiltThe primary tilt plane can be in either the AP or ML direc
ComponentsLeica Angle Two™ User Manual Rev A02 © Leica Biosystems Richmond Inc. 2010 182. Unlock the rotation clamp and rotate the base of the manipu
ComponentsLeica Angle Two™ User Manual Rev A02 © Leica Biosystems Richmond Inc. 2010 19Swing-Out Use swing-out to change the primary tilt plane by 90
Leica Angle Two™ User Manual Rev A02 © Leica Biosystems Richmond Inc. 2010 2Intended Use StatementThe Angle Two™ system is designed for stereotaxic s
ComponentsLeica Angle Two™ User Manual Rev A02 © Leica Biosystems Richmond Inc. 2010 203.4 Probe ShaftA clamp attaches the probe shaft to the manipul
ComponentsLeica Angle Two™ User Manual Rev A02 © Leica Biosystems Richmond Inc. 2010 21Figure 20. The probe shaft screwAttaching a ProbeAt the botto
ComponentsLeica Angle Two™ User Manual Rev A02 © Leica Biosystems Richmond Inc. 2010 223.5 Vernier ScalesAll Leica stereotaxic instruments have verni
ComponentsLeica Angle Two™ User Manual Rev A02 © Leica Biosystems Richmond Inc. 2010 233.6 Computer The computer consists of a desktop case, monitor,
Leica Angle Two™ User Manual Rev A02 © Leica Biosystems Richmond Inc. 2010 244Software Overview This chapter describes the Angle Two™ software and it
Software OverviewLeica Angle Two™ User Manual Rev A02 © Leica Biosystems Richmond Inc. 2010 254.1 AtlasThe instrument comes with either one mouse or
Software OverviewLeica Angle Two™ User Manual Rev A02 © Leica Biosystems Richmond Inc. 2010 264.2 ModesFour modes are used during surgery: Set Target
Software OverviewLeica Angle Two™ User Manual Rev A02 © Leica Biosystems Richmond Inc. 2010 273. Manually enter the coordinates into the Target (Red)
Software OverviewLeica Angle Two™ User Manual Rev A02 © Leica Biosystems Richmond Inc. 2010 28AtlasThe Atlas mode shows the exact position of the pro
Software OverviewLeica Angle Two™ User Manual Rev A02 © Leica Biosystems Richmond Inc. 2010 294.3 Initial SettingsBefore locating the target site, yo
Leica Angle Two™ User Manual Rev A02 © Leica Biosystems Richmond Inc. 2010 3ContentsLeica Angle Two™ . . . . . . . . . . . . . . . . . . . . . . . .
Software OverviewLeica Angle Two™ User Manual Rev A02 © Leica Biosystems Richmond Inc. 2010 30Enter OffsetsOffsets are used when performing surgery o
Software OverviewLeica Angle Two™ User Manual Rev A02 © Leica Biosystems Richmond Inc. 2010 314.4 CoordinatesCoordinates can be saved or exported at
Software OverviewLeica Angle Two™ User Manual Rev A02 © Leica Biosystems Richmond Inc. 2010 321. Enter a name and description.2. In Choose Source Poi
Leica Angle Two™ User Manual Rev A02 © Leica Biosystems Richmond Inc. 2010 335AlignmentAlign the instrument whenever the software is started.Alignmen
AlignmentLeica Angle Two™ User Manual Rev A02 © Leica Biosystems Richmond Inc. 2010 345.2 ProcedureThe primary tilt of the instrument can be in eithe
AlignmentLeica Angle Two™ User Manual Rev A02 © Leica Biosystems Richmond Inc. 2010 355. Attach the clamp adaptor to the dial indicator by sliding th
AlignmentLeica Angle Two™ User Manual Rev A02 © Leica Biosystems Richmond Inc. 2010 368. Attach the other end of the alignment shaft into the manipul
AlignmentLeica Angle Two™ User Manual Rev A02 © Leica Biosystems Richmond Inc. 2010 37Horizontal AlignmentThe ML horizontal path of the probe is cont
AlignmentLeica Angle Two™ User Manual Rev A02 © Leica Biosystems Richmond Inc. 2010 3812. Use the ML drive to move the spring horizontally along the
AlignmentLeica Angle Two™ User Manual Rev A02 © Leica Biosystems Richmond Inc. 2010 39AP Tilt13. If the tilt is in the AP direction use the ML drive
AlignmentLeica Angle Two™ User Manual Rev A02 © Leica Biosystems Richmond Inc. 2010 4015. When the AP tilt is aligned, select ll To AP on the compute
AlignmentLeica Angle Two™ User Manual Rev A02 © Leica Biosystems Richmond Inc. 2010 4114. Move the spring to the bottom and then up using the DV driv
Leica Angle Two™ User Manual Rev A02 © Leica Biosystems Richmond Inc. 2010 426Using the InstrumentThis chapter describes the whole process of using t
Using the InstrumentLeica Angle Two™ User Manual Rev A02 © Leica Biosystems Richmond Inc. 2010 43Preparation1. Ensure that the instrument is clean (s
Using the InstrumentLeica Angle Two™ User Manual Rev A02 © Leica Biosystems Richmond Inc. 2010 4411. Support the head and insert the right ear bar in
Using the InstrumentLeica Angle Two™ User Manual Rev A02 © Leica Biosystems Richmond Inc. 2010 45Locating the Target Site22. Click To Target (see “To
Using the InstrumentLeica Angle Two™ User Manual Rev A02 © Leica Biosystems Richmond Inc. 2010 46Probe Placement30. Unlock the swing-out clamp and re
Leica Angle Two™ User Manual Rev A02 © Leica Biosystems Richmond Inc. 2010 477Cleaning and MaintenanceCaution Do not autoclave. The bushings and prob
Leica Angle Two™ User Manual Rev A02 © Leica Biosystems Richmond Inc. 2010 488SpecificationsInstrument DimensionsResolution, Linear ScalesDrive Advan
SpecificationsLeica Angle Two™ User Manual Rev A02 © Leica Biosystems Richmond Inc. 2010 49Tooth Bar Movement RangesProbe HolderComputerMonitorDV 20
Leica Angle Two™ User Manual Rev A02 © Leica Biosystems Richmond Inc. 2010 51IntroductionThe Leica Angle Two™ stereotaxic system is designed to perfo
Software Licence TermsLeica Angle Two™ User Manual Rev A02 © Leica Biosystems Richmond Inc. 2010 50Software Licence Terms The following License Terms
Leica Angle Two™ User Manual Rev A02 © Leica Biosystems Richmond Inc. 2010 62Instrument InstallationThe Leica Angle Two™ system needs to be assembled
Instrument InstallationLeica Angle Two™ User Manual Rev A02 © Leica Biosystems Richmond Inc. 2010 72.2 AssemblyWhen unpacked the Angle Two instrument
Instrument InstallationLeica Angle Two™ User Manual Rev A02 © Leica Biosystems Richmond Inc. 2010 82. Connect the five instrument cables into the por
Leica Angle Two™ User Manual Rev A02 © Leica Biosystems Richmond Inc. 2010 93ComponentsThis chapter describes the parts of the Leica Angle Two™ syste