MeasuringUser Manual012 53678910114
2. Measuring pitch5mm : 100Stage micrometerwith 0.1mm graduation10Measuring – User ManualCalibrating1.37mm on the stage micrometer (X)100 intervals on
11Measuring – User Manual3. Mesh pitch100 × 1mm2Calibrating6.2mm on the stage micrometer (X)10 intervals on the eyepiece graticule (Y)6.210mm = 0.62mm
Illustrations, descriptions and technical data are not binding and may be changed without notice.M2-275-0en • © Leica Microsystems (Switzerland) Ltd •
Measuring – User Manual2• Stage micrometer (1) for calibration• Graticules with various measuring pitches (2) in mm and inches• Graticule with mesh (3
3Measuring – User ManualAvoid corrosive cleaning agents and materials whichwill scratch the surfaces.Observe the information relating to care in the o
Co-ordinate the eyepieces precisely to your eyes asdescribed until you can clearly see the graticule andspecimen at the same time (free from parallax)
5Measuring – User Manual Select the lowestmagnification. Observe specimen through theeyepiece with graticule andfocus. Select the highestmagnificat
6Measuring – User ManualThe graticule is only magnified by the eyepiece.The total magnification of the specimen does howeverdepend on the lens, magnif
7Measuring – User Manual Place stage micrometer onthe table insert and setsharpness. Select the magnificationwith which you will laterwant to take m
8Measuring – User Manual Remove the stage micrometer and focus on a specimen. Count how many intervals on the eyepiece graticulecover the measuremen
9Measuring – User Manual1. Measuring pitch12mm : 120Stage micrometerwith 0.1mmgraduation ExamplesCalibrating7.8mm on the stage micrometer (X)120 inter